COVID-19 Update - This is to inform you that the Government of India has announced a complete lockdown in India 22nd March 2020 to 3rd May 2020. As a result, our offices will now be closed till 3rd May 2020 and all our employees will be working from home. Office telephones will not be answered, and therefore you are requested to direct all your queries related to manuscript submission, review process, publication etc. at below mentioned details. editor@innovativepublication.com, rakesh.its@gmail.com, Mob. 8826373757, 8826859373, 9910947804


Print ISSN:-2394-6369

Online ISSN:-2394-6377

CODEN : IJCBK6

Current Issue

Year 2020

Volume: 7 , Issue: 1

Article Access statistics

Viewed: 778

Emailed: 0

PDF Downloaded: 233

International Journal of Clinical Biochemistry and Research


Comparison of glucose levels using glucometer and GOD-POD Method in diabetic patients


Full Text PDF Share on Facebook Share on Twitter


Author Details : Sherie Pankaj Sharma, Ashish Prakash Anjankar, Anita Kale

Volume : 4, Issue : 1, Year : 2017

Article Page : 6-10


Suggest article by email


Abstract

Introduction: Glucometer is widely used at hospitals and homes as a first line tool to get an idea about the current blood glucose levels. Since glucometers are used for making important decisions, it is essential that their accuracy should be comparable to those of standard laboratory analyzers. So the present study was aimed to compare and correlate the glucose levels using glucometer and GOD-POD method in diabetic patients.
Methods: 60 diagnosed type 2 diabetes mellitus patients were selected for the study after institutional ethical committee’s clearance was obtained. These were divided into three groups based on their plasma glucose levels by GOD-POD method.
Group 1: 20 Patients having plasma glucose levels ≤ 110 mg/dl.
Group 2: 20 Patients having plasma glucose levels between 111 mg/dl to 250 mg/dl.
Group 3: 20 Patients having plasma glucose levels ≥ 251 mg/dl.
In this study, we measured plasma glucose levels of each patient simultaneously with glucometer and GOD-POD method (semi-automatic analyzer) in the clinical laboratory.
Results: In our study, we found that that at very high glucose readings (i.e. in group 3 patients having plasma glucose levels ≥ 251 mg/dl), glucometer overestimates glucose results. So, these values do not accurately reflect actual plasma glucose levels.
Conclusions: Very high glucose values obtained using glucometers should be cautiously interpreted and verified with centralized laboratory. The current practice of performing only single readings with glucometers can lead to misdiagnosis. Medical professionals should depict diabetic patients the importance of periodic centralized laboratory glucose testing.

Keywords: Diabetes, Glucose levels, Glucometer, GOD-POD method

How to cite : Sharma S P, Anjankar A P, Kale A, Comparison of glucose levels using glucometer and GOD-POD Method in diabetic patients. Int J Clin Biochem Res 2017;4(1):6-10