COVID-19 Update - This is to inform you that the Government of India has announced a complete lockdown in India 22nd March 2020 to 14th April 2020. As a result, our offices will now be closed till 14th April 2020 and all our employees will be working from home. Office telephones will not be answered, and therefore you are requested to direct all your queries related to manuscript submission, review process, publication etc. at below mentioned details. editor@innovativepublication.com, rakesh.its@gmail.com, Mob. 8826373757, 8826859373, 9910947804

Article view: 391

Article download: 130

International Journal of Clinical Biochemistry and Research


Use of deproteinizer to eliminate interference of high bilirubin on measurement of serum creatinine by the kinetic jaffe reaction


Full Text PDF Share on Facebook Share on Twitter


Author Details : Vijaysinh Parmar, Hiren Sanghani, Asha Khubchandani

Volume : 5, Issue : 2, Year : 2018

Article Page : 226-229


Suggest article by email


Abstract

Introduction: Serum creatinine is a useful diagnostic parameter of kidney function. Its concentration reflects glomerular filtration rate. Because of negative interference of bilirubin over creatinine level, may mislead the clinician. The most common and serious problems in the determination of serum creatinine concentration by the kinetic Jaffe method is the negative interference by bilirubin.
Materials and Methods: This cross sectional study was done on 160 patients which have high bilirubin level, taken from B. J. Medical College, Civil hospital, Ahmedabad from August 2013 to December 2013. These samples were studied for interference of total bilirubin on measurement of Serum creatinine level before and after treating with deproteinizer (trichloroacetic acid 0.55 mmol/ltr). Serum bilirubin and serum creatinine level were measured by modified Jendrassik & Groff’s method and kinetic Jaffe’s reaction respectively by using Erba XL-640 fully auto analyzer.
Results: In this study significant difference was found in Serum Creatinine level with the samples of high bilirubin content before and after treating with deproteinizer (trichloroacetic acid) (p<0>Preprecipitation of albumin along with bilirubin by trichloroacetic acid beforehand is suggested as an approach to correct the bilirubin interference.
Conclusion: From this study it is concluded that bilirubin have negative interference on serum creatinine level. And this is corrected by using deproteinizer trichloroacetic acid.

Keywords: Serum Creatinine, Serum bilirubin, Negative .intereference

Doi : 10.18231/2394-6377.2018.0047

How to cite : Parmar V, Sanghani H, Khubchandani A, Use of deproteinizer to eliminate interference of high bilirubin on measurement of serum creatinine by the kinetic jaffe reaction. Int J Clin Biochem Res 2018;5(2):226-229

Copyright © 2018 by author(s) and Int J Clin Biochem Res. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (CC BY 4.0) (creativecommons.org)